Check the lamp brightness.
The lamp brightness can be varied (see Page 107).
Brightness adjustment has an effect on the surgical field illumination.
•
Check the lamps
Primary lamp and backup lamp are both intact.
•
Check the stand for secure positioning
The locking tabs of at least two casters have been pressed and prevent
the stand from starting to move by itself (see Page 81).
•
Check the user settings.
The correct user profile has been selected and the settings have been
checked (see Page 103).
•
Check the unobstructed movement of the system.
The system has been positioned in such a way that it can be moved
away from the patient at any time.
•
Check foot control panel
Suspension system and surgical microscope must perform the specified
functions (see Page 110).
NOTE
Risk of injury when tilting the surgical microscope!
If the surgical microscope is tilted heavily, the co-observation tube may swing
downward unintentionally, or an accessory attached to the surgical microscope
may swing in unintentionally.
This can cause contact with and injury to the patient.
•
Always use extreme caution when adjusting the surgical microscope.
–
Hold onto the surgical microscope firmly.
–
Loosen the friction knob only a little.
–
To the extent possible, tilt the surgical microscope only by a small
angle range.
–
Finally, tighten the friction knob firmly again.
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...