OPMI Lumera i
Preparation for use
Balancing the system
Do not perform balance setting until the surgical microscope and all accessories
have been mounted!
•
Move the suspension arm (3) into its horizontal position and tightly hold
it with one hand.
•
Loosen knob (2) for adjusting the friction of the suspension arm’s upward/
downward movement.
•
Move suspension arm (3) slightly up and down. At the same time, turn
adjustment screw (1) until you think that the spring force is sufficient to
compensate for the weight of the surgical microscope and accessories.
Turn clockwise to increase the spring force, turn counterclockwise to
reduce it.
•
Continue to perform this adjustment until the surgical microscope moves
neither upward nor downward. The surgical microscope is fully balanced
when it remains stationary in all positions within the work area.
Fig. 26: Balancing the system
CAUTION
Risk of crushing – mind your fingers!
Fingers may be crushed between the carrier and suspension arms.
•
Never touch the area between the suspension and carrier arms while you
are performing the balancing procedure or moving the suspension arm.
2
1
3
Summary of Contents for OPMI Lumera i
Page 4: ...OPMI Lumera i Version 8 0 Page 4 G 30 1720 en...
Page 32: ...Safety measures OPMI Lumera i Version 8 0 Page 32 G 30 1720 en...
Page 35: ...Version 8 0 G 30 1720 en Page 35 OPMI Lumera i Device overview Fig 4 System overview 3 1 2...
Page 61: ...Version 8 0 G 30 1720 en Page 61 OPMI Lumera i Preparation for use...
Page 83: ...Version 8 0 G 30 1720 en Page 83 OPMI Lumera i Preparation for use...
Page 88: ...Preparation for use OPMI Lumera i Version 8 0 Page 88 G 30 1720 en...
Page 97: ...Version 8 0 G 30 1720 en Page 97 OPMI Lumera i Operation...
Page 99: ...Version 8 0 G 30 1720 en Page 99 OPMI Lumera i Operation Fig 35 Menu structure 2 3 7 6 5 4 1 8...
Page 182: ...OPMI Lumera i Version 8 0 Page 182 G 30 1720 en...
Page 183: ...Version 8 0 G 30 1720 en Page 183 OPMI Lumera i Blank page for your notes...