ZEISS
10 Technical Data and Conformity | 10.2 Installation Requirements
Parameter
Requirement
Corners
Min. 1.5 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m
2
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 3.5 m × 5.0 m × 2.3 m
Service area
Min. 1.0 m at each side
Entrance
Min. 0.8 m wide
Hallways
Min. 1.0 m wide
Corners
Min. 1.2 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m
2
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 3.6 m × 5.0 m × 2.3 m
Service area
Min. 0.8 m at each side
Entrance
Min. 0.8 m wide
Hallways
Min. 1.0 m wide
Corners
Min. 1.2 m wide
Transport ways
Free of staircases
Installation cate-
gory
II
Floor stability
> 1000 kg/m
2
Parameter
Requirement
Installation site
Exclusively inside buildings
Recommended
room size
Min. 3.6 m × 4.0 m × 2.3 m
Service area
Min. 0.8 m at each side
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000
115