ZEISS
8 Troubleshooting | 8.6 Detectors
8.6 Detectors
8.6.1 Lubricating the Rod
The rod from the aSTEM and BSD4 detector mechanics needs to be lubricated once a year with
TEM oil 300.
Parts and Tools
Designation
Part no.
000000-0484-955
Isopropanol
–
Lint-free cloth
–
Lint-free gloves
–
CAUTION
TEM oil 300 may be irritating to skin and eyes.
4
Avoid contact with skin.
4
Wear suitable gloves.
4
In case of contact with the eyes, rinse immediately with plenty of water and seek medical
advice.
4
After contact with the skin, wash immediately with plenty of water and soap.
CAUTION
Risk of injury: Isopropanol
Isopropanol is highly flammable and irritating to the eyes.
Vapors may cause drowsiness and dizziness.
4
Wear suitable gloves.
4
Keep away from sources of ignition.
4
Do not smoke.
4
In case of contact with the eyes, rinse immediately with plenty of water and seek medical
advice.
4
Avoid contact with skin.
4
Do not breathe vapor.
NOTICE
Risk of property damage: Unsuitable lubricants
When using unsuitable lubricants, the vacuum system may be contaminated.
4
Only use TEM oil 300 for lubricating.
1. Retract the respective detector.
2. Clean the rod with isopropanol with a clean, lint-free cloth.
3. Spread some drops of TEM oil 300 across the rod.
Use a clean, lint-free cloth.
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000