8 Troubleshooting | 8.5 Power Circuit
ZEISS
7. Click Cal I Probe.
8. To confirm, click Yes.
9. To store the calibration, click Save.
10. Deactivate the Spot checkbox.
8.5 Power Circuit
8.5.1 Checking the Position of the Circuit Breakers
Safety Information
NOTICE
Risk of property damage: Persisting electrical problems
Tripped circuit breakers may be a hint for an electrical problem in the microscope.
4
If a circuit breaker keeps tripping, de-energize the microscope completely and contact your
ZEISS service representative for assistance.
No.
Value
Circuit
F1
10 A
Power supply unit
F2
10 A
F3
10 A
Airlock, pre-vacuum pump
F4
10 A
Internal heaters
1. Check if one of the circuit breakers on the rear
side of the plinth is tripped.
2. If one of the circuit breakers is tripped, push it upwards.
104
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000