ZEISS
6 Operation | 6.1 Starting the System
6 Operation
6.1 Starting the System
6.1.1 Energizing the Microscope
WARNING
Risk of injury: Restart after emergency off
If the reason for the emergency off is not eliminated, it may be dangerous to restart the micro-
scope.
4
If the microscope has been de-energized due to an emergency, ensure that the reason for
the emergency off does not exist anymore.
4
Make sure it is safe to energize the microscope.
Prerequisite
ü
The microscope has been de-energized by turning the
MAIN
switch to the
OFF
position.
ü
Alternatively, the microscope has been de-energized by pressing the
Emergency Off (EMO)
button
.
1. Verify that the main shut-off valves for compressed air, nitrogen, and cooling water at the
facility installation are operable.
Otherwise open and unlock the main shut-off valves.
2. If the Emergency Off (EMO) option is installed, pull the
EMO
button to release it.
3. Set the
MAIN
switch to the
ON
position.
4. If the Emergency Off (EMO) option is installed,
press the green
S2
button to confirm the set-
ting of the
MAIN
switch.
Instruction Manual ZEISS Crossbeam 550L, Crossbeam 550 | en-US | Rev. 3 | 349500-8122-000
65
Summary of Contents for Crossbeam 550
Page 135: ......