Zeiss LSM800-02 + Airyscan confocal microscope  

   

20/11/2019                  

System description: 

I.

 

Upright confocal microscope ZEN blue 2.6 HotFix-8 

II.

 

2 GaAsP +1 AiryScan detector (GaAsP) 

III.

 

Motorized stage for multiple position imaging  

IV.

 

Heating chamber 

V.

 

NO CO

mixer 

VI.

 

Multi-slide holder, dish holder, single slide holder 

Objectives:  

-

  Position1: Objective Plan-Apochromat 

10x / NA 0.45 Air 

(WD=2.1mm)

 

-

  Position2: 

Objective  Plan-Apochromat 

20x / NA 1.0 W

 DIC (WD=1.8mm)

 

-

  Position3: Objective Plan-Apochromat 

20x / NA 0.8 Air

 

-

  Position4: Objective 

Plan-Apochromat 

40x/1.4 Oil DIC 

(WD=0.13 mm) 

 

 
 

-

  Objective Plan-Apochromat 

63x / NA 1.4 Oil 

(not mounted)

 

-

  Objective Plan-Apochromat 40x / NA 1.2 Water  (not mounted)

 

 

 

Laser lines: 405, 488, 561, 640 nm 
 

Detectors: 

 

Two very sensitive Gallium arsenide phosphide photomultiplier tube (GaAsP PMTs) detectors 
with free choice of the spectral range. (Uses a tunable short pass and band pass filter in front 
of the PMTs) 

 

Be very careful with using it to prevent instrument damage! If oversaturated, detectors will 
switch off automatically. 

 

1 GaASP 

PMT for the 

Airyscan

 detector that can function as a normal detector in confocal 

mode (detection range adjustable by a tunable short pass and a tunable long pass filter, no 
bandpass filters

 

An ESID detector for transmitted light. 

Simultaneous detection of 3 dyes possible. 

 

VII.

 

Starting up: 

 

1.

 

Switch on the multiple socket (1) “

Hanging on the right side of the rack

”. Provides power 

to the computer and microscope controllers 

Summary of Contents for LSM800-02

Page 1: ...an Apochromat 40x NA 1 2 Water not mounted Laser lines 405 488 561 640 nm Detectors Two very sensitive Gallium arsenide phosphide photomultiplier tube GaAsP PMTs detectors with free choice of the spec...

Page 2: ...the main microscope menu 3 Since this is an upright system changing objectives can only be manually done by rotating the objective revolver Rotation should work very easily when objectives are in the...

Page 3: ...ed dyes are not set to the 1st detector due to the Variable dichroic arrangement If you are using the detector for transmitted light ESID activate ESID represented in the light path scheme Adjust inte...

Page 4: ...ber of pixels and the pixel size according to Nyquist sampling criteria Note More pixel than suggested by the software as optimal rarely make sense On the other hand feel free to decrease the pixel nu...

Page 5: ...s tab Important to make sure that the software images the positions in the order you added them make sure that you DISABLE the feature tile regions positions in the options menu of the Tiles bar see p...

Page 6: ...rked with a home logo as this setting will be used in the next startup of the system and might cause potential damage to GaAsP PMTs 2 Close the ZEN Software 3 Shut down the computer 4 Turn off the coo...

Page 7: ...ser should be avoided at all times Laser safety goggles are situated at all workspaces and should be used in any situation where potential contact of eyes with Laser light of the classes 3B or 4 is po...

Reviews: