background image

 

 

Set pinhole to Airy Unit (AU) 1 in the channel dialog to achieve confocality. Note, that activating 
the ‘1AU’ button by clicking on it once (gets light blue background color) fixes the pinhole to 
1AU. In this case, the match pinhole function in the Z-stack setup will not change pinhole sizes. 

 

In the ‘Channels’ dialog you can adjust the pinhole, gain and laser power manually 

 

Press the “Auto exposure” button and the software will adjust the gain of the PMTs but not the 
laser power, according to the brightness of your dyes. 

 

What Auto exposure does, is to increase the PMT voltages until about 1% of all pixels are 
overexposed. If the resulting gain is below 400 V, decrease the laser power. If the gain is above 
900 V, increase the laser power. 

 

To zoom in your image, go to ‘Acquisition mode’ and adjust the slider accordingly. With “Crop”, 
you can select the desired area of your image, then press “Live” or “Snap” to apply. “Reset all” 
will remove the zoom again. Depending on the scan speed, the lowest zoom is 0.5 and not 1.0. 

 

Before taking an image, press ‘optimal’ in the ‘acquisition mode’ tab to adjust optimal number 
of pixels and the pixel size according to Nyquist sampling criteria. 
Note: More pixel than suggested by the software as ‘optimal’ rarely make sense. On the other 
hand, feel free to decrease the pixel number to increase acquisition speed at the cost of 
resolution. 

 

To capture single images, click on ‘Snap’ (Snapped images are overwritten if you press 

Live

 

before saving them) 

 
 

1.

 

Setting up Z-stack:   
 

 

Enable Z-Stack and the Z-Stack dialog will appear 

 

Press “Live” and set your first and last slice 

 

The software will suggest optimal slice number and interval which you can change manually 

 

Activate the 

 ‘Optimize sectioning and step’ checkbox to see slice overlapping. Here you can 

again set an optimal interval to ensure ~50% overlap between sections. 

 

If you wish to use the match pinhole function, make sure that the ‘1AU’ button in the 
‘Channels’ menu is deactivated on all tracks (no blue background of button). This function will 
match pinhole sizes to the optimal pinhole size of the lowest wavelength of your imaging setup 
and ensure that all sections have the same thickness. 

 

Press “Start experiment” 
 

2.

 

Setting up time series:  
 

 

Enable “time series” 

 

Time series dialog will open 

 

Set cycle number and interval time 

 

Press “start experiment” 

 

3.

 

Tile Scan (in this mode you have: Tiles, Multipositions): 
 

 

Enable “tile scan” 

Summary of Contents for LSM800-02

Page 1: ...an Apochromat 40x NA 1 2 Water not mounted Laser lines 405 488 561 640 nm Detectors Two very sensitive Gallium arsenide phosphide photomultiplier tube GaAsP PMTs detectors with free choice of the spec...

Page 2: ...the main microscope menu 3 Since this is an upright system changing objectives can only be manually done by rotating the objective revolver Rotation should work very easily when objectives are in the...

Page 3: ...ed dyes are not set to the 1st detector due to the Variable dichroic arrangement If you are using the detector for transmitted light ESID activate ESID represented in the light path scheme Adjust inte...

Page 4: ...ber of pixels and the pixel size according to Nyquist sampling criteria Note More pixel than suggested by the software as optimal rarely make sense On the other hand feel free to decrease the pixel nu...

Page 5: ...s tab Important to make sure that the software images the positions in the order you added them make sure that you DISABLE the feature tile regions positions in the options menu of the Tiles bar see p...

Page 6: ...rked with a home logo as this setting will be used in the next startup of the system and might cause potential damage to GaAsP PMTs 2 Close the ZEN Software 3 Shut down the computer 4 Turn off the coo...

Page 7: ...ser should be avoided at all times Laser safety goggles are situated at all workspaces and should be used in any situation where potential contact of eyes with Laser light of the classes 3B or 4 is po...

Reviews: