5
•
Installation
Page 74
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
No.
Description
5b
Isocratic pump, quaternary pump
: Wash line from pump head to seal
wash detector
Binary pump, dual pump
: Wash line from left pump head to right pump
head
5c
Binary pump, dual pump
: Wash line from right pump head to seal wash
detector
6
Seal wash detector (drop detector)
Follow these steps
Setting up the seal wash system comprises the following steps:
1. Engaging the peristaltic tubing
2. Connecting the seal wash line between autosampler and pump
3. Routing the seal wash line to the solvent rack
4. Connecting the seal wash line to the seal wash reservoir
5. Flushing the seal wash system
Follow the steps in the respective sections further down.
Engaging the Peristaltic Tubing
In the pump, engage the peristaltic tubing (PharMed™ tubing) in the seal
wash pump:
Figure 23: Seal wash pump
No.
Description
1
Pump lever
2
Pump rotor
3
Tubing holder (peristaltic tubing)
4
Tubing connector
5
Seal wash line from autosampler
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...