7.11 Testing the Pump for Leakage (General Test)
When
After maintenance on the fluidic system of the pump, for example on
the flow connections, pump head, or check valves
TIP
With the dual pump, test each pump separately.
Parts and additional items required
•
Fitting plug, Viper
•
Solvent suitable for your application
Follow these steps
1. Disconnect the capillary on the pump outlet.
2. Close the pump outlet with the fitting plug.
3. In the Chromeleon software, set the upper pressure limit to the
appropriate value:
Pump
Upper Pressure Limit
VC‑pump
60 MPa
VF‑pump
80 MPa
4. Apply a flow of 50 µL/min and use the solvent for the application.
5. Observe the pressure. When the pressure reading is as indicated in
the table, change the flow rate to a lower value.
Pump
Pressure
VC‑pump
Between 30 MPa and 40 MPa
VF‑pump
Between 60 MPa and 70 MPa
6. Have the pump deliver until the following pressure has built up:
Pump
Pressure
VC‑pump
55 MPa
VF‑pump
75 MPa
7. When the pressure has built up, apply the following flow:
Pump
Flow
VC‑P10, VC‑P20,
VC‑P32, VC‑P40
1 µL/min
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...