7
•
Maintenance and Service
Page 168
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
7.8
Mixing System
TIP
The information and references to the static mixer in this section
apply equally to the inline filter if installed.
Follow the sequence of steps for the maintenance procedure that you
want to perform:
•
To test the static mixer for permeability, see
.
•
To replace the static mixer or capillary mixer, see
Static Mixer or Capillary Mixer (
7.8.1
Testing the Static Mixer for Permeability
TIP
With the dual pump, test each static mixer separately.
When
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
Additional items required
•
Water as solvent
•
Tissue
Follow these steps
NOTICE
Turning the static mixer in the mounting bracket when loosening or
tightening capillaries can damage the mounting bracket.
To prevent the mixer from turning in the bracket, hold the mixer with an
open-end wrench when you loosen or tighten capillaries on the mixer.
1. Disconnect the capillary on the outlet of the static mixer.
2. To absorb liquid leaving the static mixer, place the tissue on the
mixer outlet.
3. Apply a flow of 2 mL/min using water.
4. Observe the pump pressure.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...