9
•
Pump-Specific Information
Page 232
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
9.3.4
Setting the Solvent Composition
For increased flexibility, the pump is shipped with two 3-port solvent
selectors. They allow forming binary gradients from three solvents per
channel. The solvent to be used for each channel is selected in the user
interface.
Follow these steps
1. In the Instrument Configuration Manager, in the dialog box for the
pump, check and the change the following if required:
¨
Number of the solvents to be used with the pump
¨
Standard solvent names
These solvent names appear in the Chromeleon user interface.
2. In the Chromeleon Client, set the solvent composition. For each of
the two channels, select the solvent and set the percentage of
solvent that you want to deliver.
Selecting the solvent
1. For channel A, set
%A_Selector
to
%A1
or
%A2
or
%A3
.
2. For channel B, set
%B_Selector
to
%B1
or
%B2
or
%B3
.
Determining the percentage of solvent to deliver
Set the percentage for solvent B (
%B
). Solvent A automatically delivers
the remaining volume.
Examples
•
To deliver 100% of solvent B2:
Set
%B_Selector
to
%B2
and set
%B
to 100%.
•
To deliver 100% of solvent A3:
Set
%A_Selector
to
%A3
and set
%B
to 0%.
•
To deliver 30% of solvent A1 and 70% of solvent B3:
Set
%A_Selector
to
%A1
and
%B_Selector
to
%B3
, and set
%B
to
70%.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...