Troubleshooting
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 203
Message and Code
Description and Remedial Action
Code 4100 Purge pressure limit exceeded
A pressure of more than 5 MPa built up after a purge cycle has
started.
Perform the following sequence of steps:
1. Open the purge valve if it is not yet open.
2. Check the flow path for indications of blockage. See also
Code 4098 Upper pressure limit exceeded
.
3. Take appropriate remedial action as necessary.
4. Mute the alarm by pressing the MUTE ALARM
button on the
pump keypad.
5. Start a new purge cycle.
Code 4101 Not ready. Please run self-
test.
The pump is not ready because the self-test was not successful.
Perform the following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4112 Self-test failed because of
fluctuating pressure.
Changes in pressure have occurred during the self-test. Perform
the following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4113 Self-test failed because
pressure is out of range.
The pressure has not been down to zero during the self-test.
Verify that the pressure sensor cables are properly connected.
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
If the message appears again, the system pressure sensor may be
defective. Contact Technical Support.
Code 4114 The device is busy. Please
retry after 1 minute.
The pump is not ready because, for example, a self-test is running.
Retry after 1 minute. If the message appears again, perform the
following sequence of steps:
1. Open the purge valve.
2. Repeat the self-test.
3. Close the purge valve.
Code 4115 Working piston pressure zero
out of range.
The message may appear for the pressure sensor in a pump head
or for the system pressure sensor:
•
If the message appears for only the right pump block or the left
pump block, the cable for the pressure sensor in the pump head
may not seat properly in the
P-WORK
connector.
•
If the message appears for both pump blocks at the system time,
the cable for the system pressure sensor may not seat properly
in the
P-SYS
connector.
Verify that the pressure sensor cable is properly plugged into the
connection port and retry. If the problem persists, contact
Technical Support.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...