7.10 Replacing the Purge Valve Knob
When
•
Leakage is observed around the valve knob when the valve is open.
•
Leakage is observed on the purge outlet when the valve is closed.
Parts required
Purge valve knob (with integrated seals)
Figure 59: Purge valve knob with integrated seals
No.
Description
1
Purge valve knob
2
Ring seal
3
Cap seal
NOTICE
Avoid holding the new valve knob by the seal end. Scratches on the seals
will prevent a proper seal and cause leakage.
Additional items required
•
Solvent suitable for purging to flush out harmful substances
•
Cleaning swab
•
Solvent suitable for your application
Preparations
1. To flush out harmful substances, purge the pump with an
appropriate solvent.
2. Turn off the pump flow and wait until the system pressure is down
to zero.