10
•
Modifying the Pump for Specific Applications
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 245
8.
With Chromeleon 7.2.8 and later:
In the Chromeleon software, set
the
CapillaryMixer
property to the value indicated on the capillary
mixer.
9. Apply the flow rate of your application and have the pump deliver
for a short time.
10. Inspect the flow connections on the mixer for indications of leakage
before resuming operation.
11. In the Chromeleon software, consider updating the Predictive
Performance information for the static mixer (
MixerChanged
command).
When operating the pump, observe the following
•
Operating the pump with a mixing system with a volume lower than
the volume installed in the pump upon shipment reduces the
gradient delay volume but increases the ripple.
•
The maintenance procedures are the same as for the mixing system
with which the pump was shipped (see
).
10.1.2.4
Installing the 35 µL Inline Filter Set
Parts required
•
Set inline filter, volume: 35 µL, comprising an inline filter (static
filter, volume: 10 µL) and a capillary mixer (volume: 25 µL)
•
Capillary to connect the pump to the autosampler
Tools and additional items required
•
Depending on the fitting connection on the capillaries
¨
Viper capillary: Viper knurl (usually present on the capillary)
¨
Conventional capillary (non-Viper): Open-end wrench, size 1/4"
•
Solvent suitable for purging
•
Screwdriver, Torx T10
Preparations
1.
If not yet done:
Remove the mixing system with which the pump is
2. The inline filer requires a different mounting bracket. Install the new
mounting bracket. The mounting bracket is included in the inline
filter ship kit.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...