7
•
Maintenance and Service
Page 192
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
3. Grasp the slide-in module by the leak tray and pull the module out
of the enclosure by approximately 10 cm.
Figure 63: Pulling out the slide-in module
NOTICE
The slide-in module can fall down when pulling it out of the enclosure
too far.
Pull out the slide-in module just far enough so that you can grasp it on
both sides from below.
4. Remove the slide-in module from the enclosure. The following steps
require a team effort:
a) Take the slide-in module on both sides from below.
b) Pull the slide-in module from the rails towards the front.
c) Place the slide-in module on a clean and stable surface.
7.16.2 Returning the Slide-In Module
Preparation
If not yet done:
Remove the slide-in module from the enclosure. See
Removing the Slide-In Module (
Follow these steps
1. Install the slide-in module to the enclosure of the replacement
device. Follow the steps for inserting the slide-in module in the
enclosure and tightening the captive screws in
.
2. Follow the instructions in
.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...