6
•
Operation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 93
•
The pumps are shipped with the following piston seals:
Pump
Piston Seals
VC‑pumps except
VC‑P21 and VC‑P33
The pumps are shipped with UHMW-PE piston seals. Using
tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals (part
no. 6040.0306). In this case, also install new pistons.
Exchanging the piston seals but keeping the pistons that
were used with the UHMW-PE piston seals may impair the
seal life of the PTFE seals. In addition, note that slightly
increased abrasion may be observed with these seals,
depending on the application. Consider testing the
permeability of the static mixer regularly at short intervals.
VC‑P21 and VC‑P33
pumps
The pumps are shipped with carbon-fiber filled PTFE piston
seals. Slightly increased abrasion may be observed with
these seals, depending on the application. Consider testing
the permeability of the static mixer regularly at short
intervals.
VF‑pumps
The pumps are shipped with UHMW-PE piston seals. Using
tetrahydrofuran, ketones, or ammonium hydroxide as
solvents can damage the seals.
Under harsh conditions, consider replacing the UHMW-PE
piston seals with carbon-fiber filled PTFE piston seals (part
no. 6040.0306). In this case, also install new pistons.
Exchanging the piston seals but keeping the pistons that
were used with the UHMW-PE piston seals may impair the
seal life of the PTFE seals. In addition, note the following:
•
Slightly increased abrasion may be observed with the
PTFE seals, depending on the application. Consider
testing the permeability of the static mixer regularly at
short intervals.
•
The pump should not be operated with pressures higher
than 70 MPa.
•
The pump is no longer biocompatible.
•
For best seal performance, do not recycle or recirculate solvents.
Avoid using methanol from aluminum reservoirs.
•
To reduce the growth of algae, consider using amber glassware or
using appropriate additives, for example, formic acid.
•
After operation, flush out buffers and solutions that form peroxides.
•
Never leave buffers, salt solutions, or aggressive solvents in the
system for a longer period without flow.
•
Note the special properties of the solvents, such as viscosity, boiling
point, or UV absorption.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...