8
•
Troubleshooting
Page 212
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
8.5.3
Testing the Static Mixer for Permeability
TIP
The information and references to the static mixer in this section
apply equally to the inline filter if installed.
When
If you observe pressure pulsation, inconstant pressure, or high
backpressure at the column and pump
Additional items required
•
Water as solvent
•
Tissue
Follow these steps
Test the static mixer by following the steps in
.