10
•
Modifying the Pump for Specific Applications
Page 246
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Installing the New Mounting Bracket
1. Unplug the system pressure sensor cable from the
P-SYS
connector.
2. Remove the seal wash detector assembly (funnel and detector).
Grasp and pull the assembly upward by the funnel. The detector
including the tubing and cable can remain connected to the funnel.
Figure 79: Removing the seal wash detector assembly
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line
3. To attach the new mounting bracket inside the pump (above the
purge/sensor unit):
a) Remove the push rivets that cover the mounting holes while no
mounting bracket is installed.
b) Using the screws shipped with the mounting bracket, attach the
new bracket inside the pump.
Figure 80: Installing the mounting bracket