10
•
Modifying the Pump for Specific Applications
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 243
10.1.2.3
Installing the 100 µL Mixing System
Parts required
•
Mixing system, volume: 100 µL, comprising a static mixer (volume:
75 µL) and a capillary mixer (volume: 25 µL)
•
Capillary to connect the pump to the autosampler
Tools and additional items required
•
Depending on the fitting connection on the capillaries
¨
Viper capillary: Viper knurl (usually present on the capillary)
¨
Conventional capillary (non-Viper): Open-end wrench, size 1/4"
•
Solvent suitable for purging
Preparations
1.
If not yet done:
Remove the mixing system with which the pump is
2. For easier installation, remove the seal wash detector assembly
(funnel and detector). Grasp and pull the assembly upward by the
funnel. The detector including the tubing and cable can remain
connected to the funnel.
Figure 77: Removing the seal wash detector assembly
No.
Description
1
Funnel, seal wash detector
2
Seal wash detector
3
Detector cable
4
Seal wash line