Pump-Specific Information
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 231
9.3.3
Dwell Volume of the Pump
The gradient delay volume of an HPLC system is defined as the volume
of the entire flow path between the point where the gradient is formed
and the point where the mobile phase enters the column (see
Delay Volume, Dwell Volume and Ripple (
The pump contributes to the system gradient delay volume by the dwell
volume, which is the calculated volume between the point where the
gradient is formed and the pump outlet.
Figure 71: Dwell volume of the pump
No.
Description
1
Solvent selectors
2
Solvent degasser
3+4
Pump heads
5
Purge/sensor unit
6
Mixing system with capillary mixer and static mixer
7
Pump outlet
8
Dwell volume of the pump
For the dwell volume of the pump, see