7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 181
5. Monitor if the pressure drops at the same rate as before. If the
pressure drops significantly slower, the connection was leaky.
6. When testing the pump for leakage is complete, reset the upper
pressure limit to the value used before the leakage test.