Troubleshooting
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 205
Message and Code
Description and Remedial Action
Code 4125 Degasser malfunction
The vacuum level monitoring function detected insufficient
degasser vacuum.
Turn the pump off and on again. Check the degasser vacuum in
the Chromeleon software. After about 1 minute, the setting
should change from
NotOk
to
Ok
. If the degasser vacuum is still
insufficient, the degasser may be leaky. Locate the
DegasserPressure
property in the
Command
window, write down
the pressure reading, which can provide helpful information to
identify the source for the problem, and contact Technical
Support.
Code 4127 The pump drive is still in
undock position. Please execute a dock
command.
The message alerts you during pump maintenance:
This message alerts you during pump maintenance (for example,
when replacing the piston seals or pistons) when you try to start
the pump while the pistons are not yet in the correct position for
normal operation. To return the pistons into the correct position,
perform the following sequence of steps:
1. Open the purge valve.
2. Dock the pistons.
3. Close the purge valve.
The message alerts you during internal pump maintenance:
If the message appears during the internal maintenance
procedure, perform an additional purge cycle.
Code 4130 Pressure buildup during
maintenance operation.
A pressure of more than 0.5 MPa built up during piston
maintenance or piston seal maintenance. To reduce the pressure,
perform the following sequence of steps:
1. Open the purge valve.
2. Dock or undock the pistons, depending on when the message
appeared.
3. Close the purge valve.
If the message appears during the internal maintenance
procedure, perform an additional purge cycle.
Code 4148 Can’t perform this command
while the flow is on.
You tried to run a self-test while the flow rate was not yet down
to zero. Apply a flow of 0 and repeat the command.
Code 4152 Pressure sensor malfunction.
Check the cable connection and retry.
Verify that the cable for the pressure sensor is properly plugged
into the connection port and retry. If the message appears again,
follow these steps:
1. Open the purge valve.
2. Perform a self-test.
3. Close the purge valve.
4. Restart the flow.
5. If this message appears again or if another message relating to
the pressure or a pressure sensor appears, contact Technical
Support.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...