Emptying the Silicone Tubing
NOTICE
Always empty the silicone tubing when purging is complete. When the
purge valve is closed, purge liquid remains in the tubing connected to
the purge outlet. Leaving the liquid in the tubing for a longer period
limits the tubing lifetime.
1. To absorb any liquid that might be leaving the purge outlet when the
tubing is removed, place the tissue under the outlet.
2. On the purge/sensor unit, remove the silicone tubing from the purge
outlet. The other end of the tubing remains in the purge drain port
outlet.
3. Hold the tubing so that any liquid that is present in the tubing flows
off through the drain port outlet.
4. On the purge/sensor unit, consider reconnecting the silicone tubing.