5
•
Installation
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 61
1. Insert the Viper capillary into the connection port.
2. Tighten the connection with your fingers.
3. Use the installation tool to tighten the connection further.
Figure 12: Tool on torque toothing for tightening connections
4. Tighten the connection until the torque limitation has been reached,
which is indicated by a clicking sound.
5. Check whether the connection leaks. If leakage exists, follow the
steps further down.
Resolving Leakage with Viper Fittings with Torque Toothing
1. Remove the capillary.
2. Clean the capillary ends carefully by using a lint-free tissue wetted
with isopropanol.
3. Reinstall the capillary.
4. If the connection continues to leak, install a new Viper capillary.
To loosen a connection, use this side of the installation tool.
Figure 13: Tool on torque toothing for loosening connections
Connecting
Viper Fittings with Knurls
To connect Viper capillaries with knurls, follow these steps:
NOTICE
•
Tighten or loosen Viper capillaries
only
with your fingers. Do not use
tools other than the knurl that comes with the capillary.
•
To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries
only
when the system pressure is down to zero.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...