Follow these steps
1. To remove the purge valve knob from the purge/sensor unit, turn
the valve knob counterclockwise all the way and pull it straight off
the purge/sensor unit.
2. Before installing the new purge valve knob, clean the opening in the
purge/sensor unit with a cleaning swab.
3. Insert the new purge valve knob into the purge/sensor unit.
4. To tighten, turn the valve knob clockwise with your fingers until it is
finger-tight.
NOTICE
Turn the valve knob only with your fingers. Overtightening may damage
the purge valve seals.
5. Resume operation.
6. Inspect the purge outlet. If liquid leaves the outlet, the valve knob
may not be tightened enough. Tighten a little more.