When tightening or loosening a Viper capillary, observe the handling
instructions for the respective fitting further down.
Connecting Viper Fittings with Torque Toothing
To connect Viper fittings with torque toothing, follow these steps:
NOTICE
•
Tighten or loosen Viper capillaries
only
with your fingers. Do not use
tools other than the knurl that comes with the capillary.
•
To avoid damage to the capillary or connection, tighten and loosen
the Viper capillaries
only
when the system pressure is down to zero.
•
Mind the correct orientation of the tool when tightening or
loosening connections (see the pictures further down in this
section).
Figure 11: Viper fitting with torque toothing
No.
Description
1
Capillary
2
Torque toothing