7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 167
7.7.3
Cleaning the Check Valves
When
•
Dirty check valves
•
Once a year
Parts and additional items required
•
Syringe (12 mL), filled with isopropanol
•
Ultrasonic bath
Preparations
Remove the check valves that you want to clean. It is recommended to
clean all check valves of the pump head at the same time. To remove the
check valves, follow the related steps in
Replacing the Inlet Check Valves
Replacing the Outlet Check Valves (
as
applicable.
Follow these steps
1. In the direction of flow, fill the check valves with isopropanol, using
the syringe.
2. For 10 minutes, place the check valves in the ultrasonic bath.
3. In the direction of flow, rinse the check valves with isopropanol,
using the syringe.
4. Reinstall the check valves, following the related steps in
as applicable.
5. In the Chromeleon software, consider updating the Predictive
Performance information for the check valves.