7
•
Maintenance and Service
Pumps C, Pumps F ( VC-Pxx, VF-Pxx)
Operating Manual
Page 169
5. Depends on the pressure reading:
Pressure Reading
Action
Pressure is less than 0.8 MPa
Continue with the next step.
Pressure is 0.8 MPa or higher
Replace the static mixer.
6. Reconnect the capillary to the outlet of the static mixer.
7. Test the pump for leakage. See
.
7.8.2
Replacing the Static Mixer or Capillary Mixer
When
•
Impaired permeability of the static mixer
•
Damage of static mixer
•
Leaky capillary mixer
Parts, tools and additional items required
•
As applicable
¨
Static mixer
¨
Capillary mixer
•
Depending on the capillary connection:
¨
Viper capillary: Viper knurl (usually present on the capillary)
¨
Conventional capillary (non-Viper): Open-end wrench, size 1/4"
•
Solvent suitable for purging
Preparations
1. To flush out harmful substances, purge the pump with an
appropriate solvent.
2. To remove harmful substances from the components in the flow
path after the purge unit, have the pump deliver for a short time
with the solvent that was used for purging.
3. Turn off the pump flow and wait until the system pressure is down
to zero.
Summary of Contents for thermoscientific Vanquish Pumps C
Page 4: ...Contacting Us Page 4 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 30: ...2 Safety Page 30 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 40: ...3 Device Overview Page 40 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 196: ...7 Maintenance and Service Page 196 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 214: ...8 Troubleshooting Page 214 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...
Page 290: ...13 Appendix Page 290 Pumps C Pumps F VC Pxx VF Pxx Operating Manual ...