2
Maintaining the DSQ II
Recommended Maintenance Schedule
Thermo Scientific
DSQ II Hardware Manual
23
Recommended Maintenance Schedule
To perform maintenance on the DSQ II, you will need to disassemble, clean, and replace
particular components.
is a factory recommended maintenance schedule based on
running multiple samples eight hours a day. You can adapt the schedule according to the
cleanliness of your samples and the number of analyses you make per day. We recommend
that you keep a maintenance log of the DSQ II service so that you can identify variations in
normal operation, which will make maintenance even easier when the need arises.
Table 2.
Recommended Maintenance Schedule
Procedure
Replace
1 Month
4 or 6
Months
As
Needed*
Cleaning the Ion Source Components
(see page
), which may include:
•
Cleaning the Ion Volume with an Inlet Valve
(see page
•
Cleaning the Ion Volume Without an Inlet Valve
(see page
)
•
Cleaning the Ion Source, Lenses and Prefilter
(see page
)
•
Replacing the Ion Source Filament
(see page
Replacing the Electron Multiplier
(see page
), which may include:
Maintaining the Turbomolecular Pump
(see page
)
Maintaining the Rotary-Vane Pump
(see page
), which may include:
•
Checking the Oil Level of a Rotary-Vane Pump
•
Adding Oil to a Rotary Vane Pump
(see page
)
•
Purging Gas from the Oil in a Rotary-Vane Pump
(see page
•
Changing the Oil in a Rotary-Vane Pump
(see page
)
**
**
Maintaining Components Outside the Vacuum Manifold
(see page
*
As Needed
depends on how close the component is to the sample introduction point. For example, the Ion Volume is closer to the sample
introduction point than any other component and requires the most frequent cleaning.
** Perform every month if you use ammonia as a chemical ionization reagent gas.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...