5
Ion Source and Inlet Valve
Inlet Valve (Upgrade)
Thermo Scientific
DSQ II Hardware Manual
177
Ball Valve
The ball valve (located inside the inlet valve) is a ball with a hole through it. The ball valve is
open when the hole in the ball is aligned with the opening between the vacuum manifold and
the inlet valve chamber. When the ball valve is closed, it prevents the vacuum manifold from
venting to atmosphere.
Figure 93.
Inlet Valve Components
1
Ball Valve
6
Inlet Valve Plug
2
Inlet Valve Solenoid
7
Inlet Valve Block
3
Inlet Valve Chamber
8
Ball Valve Lever
4
Inlet Valve Seal
9
Guide Bar Housing
5
Inlet Valve Knob
10
Vacuum Manifold
1
2
3
4
5
6
7
8
9
10