4
Vacuum System and Gas Inlets
Gas Inlet Components
162
DSQ II Hardware Manual
Thermo Scientific
Gas Inlet Components
There are three gas inlets that control the input of various gases into the vacuum manifold.
The calibration gas flow module controls the calibration compound, the CI reagent gas flow
module regulates the flow of CI reagent gas into the ion source, and the vent valve solenoid
vents the vacuum manifold to the atmosphere.
Figure 82.
Gas Inlets (Left Side)
1
CI Reagent Gas Flow Module
3
Vent Valve Solenoid
2
Calibration Gas Flow Module
1
2
3
Summary of Contents for DSQ II