Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
96
DSQ II Hardware Manual
Thermo Scientific
Replacing the Lens Interface Board
To replace the lens interface, you will need a 3 mm Allen wrench.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the right and left covers, as described in "
Removing the Right Cover" on page 27
and "
Removing the Left Cover" on page 29
4. Remove the lens interface board.
Figure 55.
Disconnecting the Lens Interface Board Cables
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Cables to Low Pass Filter Board
2
Cable to Analog Board
1
2