Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
74
DSQ II Hardware Manual
Thermo Scientific
Figure 41.
Replacing the Analog Board Fuse
a.
Disconnect the analog board fuse.
b. Connect the new analog board fuse (PN 76339-0008; 2 A, very fast microfuse,
125 V). It does not matter which way it is plugged in.
5. Reattach the left, top, and front covers of the DSQ II.
6. Restart the DSQ II.
CAUTION - INSTRUMENT DAMAGE.
Only use a 125V
fuse or else you could damage the
instrument. Never replace DSQ II fuses with those of a different type, voltage, or current
rating. Also, to avoid damaging the boards, you should always wear a properly grounded
electrostatic discharge (ESD) strap on your wrist when replacing fuses.