Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
110
DSQ II Hardware Manual
Thermo Scientific
c.
Lift out the current RF generator board.
d. Disconnect the P2 cable.
5. Install the new RF generator board (PN 119590-0060).
a.
Reconnect the P2 cable.
b. Insert the new RF Generator board.
c.
Reattach the screws that hold the RF Generator board in place.
d. Reconnect the P1, J1, J2, and J3 cables.
6. Reattach the left, top, and front covers of the DSQ II.
7. Restart the DSQ II.
Replacing the RF Generator Board Fuses
To replace an RF generator board fuse, you will need:
• RF Generator board Fuse F1 (PN A0006-07608; 0.5 A, Quick Act, 5 x 20 mm, 250 V)
• RF Generator board Fuse F2 (PN A0006-07610; 1.0 A, Quick Act, 5 x 20 mm, 250 V)
• Wrench, Allen, 3 mm
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...