Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
84
DSQ II Hardware Manual
Thermo Scientific
c.
Remove the DC driver cover.
Figure 45.
DC Driver with Cover Removed
d. Disconnect the J2, J3, J4, J7, J8, and P1 cables.
e.
Remove the three screws holding the DC driver in place.
5. Install the new DC driver.
a.
Insert the new DC driver.
b. Reattach the screws.
c.
Reconnect the cables. The green cable leading to the DC driver goes on the left
terminal (J7), the black cable goes to J4, and the red cable goes to J3.
d. Reattach the DC driver cover.
e.
Reattach the left, top, and front covers of the DSQ II.
f.
Restart the DSQ II.
1
Connector to J8 Cable (to Analog
Board)
4
Three Screws Around the DC
Driver
2
Connector to J2 Cable (to RF
Detector Board)
5
Connector to P1 Cable (to Power
Module)
3
Connectors to J3, J4, J7 Cables (to
Low Pass Filter Board)
1
2
3
5
4