Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
102
DSQ II Hardware Manual
Thermo Scientific
3. Remove the top and front covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the current power module.
Figure 60.
Removing the Power Module
a.
Remove the two screws holding the power module in place.
b. Lift the power module straight up.
5. Install a new 115 V power module (PN 119655-1115) or 230 V power module
(PN 119655-1230).
a.
Insert the new power module.
b. Reattach the screws that hold the power module in place.
6. Reattach the top and front covers of the DSQ II.
7. Restart the DSQ II.
1
Power Module
(PN 119655-1115 or -1230)
2
Two Screws Around the Power Module
1
2
CAUTION - INSTRUMENT DAMAGE.
The power module voltage setting (115 V or
230 V) must match the one that was removed. Otherwise, you could damage the
instrument and rotary-vane pump. Do not attempt to change the voltage setting of the
power module. It has been configured at the factory and cannot be changed in the field.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...