Maintaining the DSQ II
Cleaning the Ion Source Components
42
DSQ II Hardware Manual
Thermo Scientific
b. Remove the ion volume from the I/R tool.
c.
Clean the ion volume, as described in
Cleaning Stainless Steel Components" on
. If you are cleaning a CI ion volume, be sure to clean out the small electron
entrance hole with a dental pick or old syringe needle. Aluminum oxide can get
trapped in the hole, which can adversely affect sensitivity.
d. Place the ion volume into the bayonet lock on the I/R tool. Make sure the alignment
arrows on the ion volume and the I/R tool are facing each other.
7. Insert the I/R tool into the DSQ II with the ion volume attached.
a.
Turn the I/R tool to the hold position
.
b. Insert the guide ball into the guide ball hole.
c.
Slide the I/R tool into the guide bar track until the guide ball is in the first groove.
d. Turn the I/R tool so the guide ball is in the first groove. This prevents the probe from
being pulled forward when the inlet valve is evacuated.
e.
Tighten the inlet valve knob to ensure a leak-tight seal.
f.
In the
Xcalibur
Insert Probe window, click the
Evacuate Inlet Valve
button.
g. Monitor the foreline pressure, which should not stay higher than 350 mTorr. If it
does, replace the inlet valve seal, as described in "
Replacing the Inlet Valve Seal" on
h. Once the inlet valve is evacuated, pull the lever up to open the inlet valve.
8. Reinstall the ion volume.
a.
Slide the I/R tool into the DSQ II with the ion volume attached until the guide ball
is slightly beyond the groove closest to the instrument.
b. Turn the I/R tool handle counter-clockwise toward the release position
.
c.
Withdraw the I/R tool about 2.5 cm (1 in.).
d. Turn the I/R tool handle to the hold position
.
e.
Slide the I/R tool toward the instrument until it feels like it cannot go in any further.
f.
Withdraw the I/R tool until the guide ball reaches the groove furthest from the
instrument.
g. Push the lever down to close the inlet valve.
h. Loosen the inlet valve knob.
i.
Remove the I/R tool.
Note
Wear clean, lint- and powder-free gloves when you handle any vacuum
manifold components.
CAUTION - INSTRUMENT DAMAGE.
The arrows on the I/R tool and ion volume must
be aligned to avoid damaging the ion source assembly.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...