Maintaining the DSQ II
Preparing the DSQ II for Maintenance
Thermo Scientific
DSQ II Hardware Manual
31
Figure 16.
Removing the Screw Holding the Left Cover in Place
5. Slide the cover toward the rear of the instrument until the tabs are at the end of the
grooved slots and then pull the cover away.
6. Reverse these steps to reinstall the left cover.
RE
V
H IG H VO LTA G E
A ND H O T SU R FA C E S
WA R N IN G
C AU TIO N
Screw holding
left cover in
place
WARNING - INSTRUMENT DAMAGE.
The analog and digital boards are attached to
interior side of the left cover, so that cover is significantly heavier than the right cover. Do
not pull the cover too far away from the instrument when you open it or you could
damage the cables leading from the boards to other parts of the instrument.