2
Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
118
DSQ II Hardware Manual
Thermo Scientific
Replacing the Vent Valve Solenoid
To replace the vent valve solenoid, you will need a lint-free cloth and an adjustable wrench.
1. Cover the work area with a lint-free cloth.
2. Shut down and vent the DSQ II, as described in
"Shutting Down the GC/MS System"
.
3. Remove the left cover, as described in "
Removing the Left Cover" on page 29
4. Remove the vent valve solenoid.
Figure 72.
Removing the Vent Valve Solenoid
a.
Disconnect the cable.
b. Unscrew the current vent valve solenoid and remove it.
c.
Unscrew the vent valve filter.
WARNING - ELECTRICAL SHOCK HAZARD.
Before proceeding. unplug the DSQ II to
avoid an electrical shock.
1
Vent Valve Solenoid
(PN 76461-1003)
3
Vent Valve Filter
2
Cable
3
2
1