1
Basic Operations
Shutting Down the GC/MS System
18
DSQ II Hardware Manual
Thermo Scientific
Shutting Down the GC/MS System
To shut down the GC/MS system:
1. Cool the TRACE GC Ultra. If you do not plan to change the column or perform
maintenance on the GC, you do not have to lower the injector temperature, but you do
need to lower the oven, injector, and transfer line temperatures to 30 °C.
2. Shut down the DSQ II.
a.
From the Instrument Setup window of
Xcalibur
, click the
Tune
button to open the
Tune window.
b. Select
Instrument | Shutdown
to begin the automatic shutdown process, which
includes:
• Displaying the
Xcalibur
Shut Down window
• Turning off the calibration and CI reagent gas
• Turning off the voltages to the ion source, quadrupole, and ion detector
• Turning off the ion source heater
• Turning off the turbomolecular pump
•
Xcalibur
counting down ten minutes for the pump to slow down
•
Xcalibur
waiting for the Ion Source to cool to < 175 °C to prevent oxidizing the
hot parts when they are exposed to air
c.
Wait for the transfer line to cool to < 175 °C.
3. Power off the DSQ II.
a.
When
Xcalibur
indicates that it is okay to turn off the main power to the DSQ II,
click
OK
.
Figure 9.
Main Circuit Breaker (Rear)
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...