2
Maintaining the DSQ II
Cleaning the Ion Source Components
54
DSQ II Hardware Manual
Thermo Scientific
3. Remove the front and top covers, as described in "
Removing the Front and Top Covers"
.
4. Remove the vacuum manifold cover, as described in "
).
5. Disconnect the 3- and 8-pin connectors from the ion source. Do not disconnect them by
pulling on the wires.
6. Loosen the two thumbscrews on the magnet yoke.
7. Slide the source spacer slightly to the left, while pulling it away from the prefilter. Once it
is free, carefully slide it to the left so that it clears the end of the transfer line. Be careful
not to damage the capillary column extending out of the transfer line.
Note
Wear clean, lint- and powder-free gloves when you handle any vacuum manifold
components.