Maintaining the DSQ II
Maintaining Components Outside the Vacuum Manifold
Thermo Scientific
DSQ II Hardware Manual
111
4. Remove the RF generator board.
Figure 67.
Removing the RF Generator Board to Replace Fuses
a.
Disconnect the P1, J1, J2, and J3 cables.
b. Remove the top two screws and loosen the bottom two screws holding the RF
generator board in place.
c.
Lift out the RF generator board.
d. Disconnect the P2 cable.
5. Replace the RF generator board fuses.
a.
Disconnect the RF generator board fuses.
b. Connect the new RF generator board fuses. It does not matter which way they are
plugged in.
1
RF Generator Board
5
Connector to J3 Cable (to Power
Module)
2
Connector to P1 Cable (to RF
Detector Board)
6
Four Screws Around the RF
Generator Board
3
Connector to J1 Cable (to Low
Pass Filter Board)
7
Connector to P2 Cable (to Analog
Board)
4
Connector to J2 Cable (to Power
Module)
7
6
1
2
3
4
5
WARNING - ELECTRICAL SHOCK HAZARD.
Only use a DSQ II fuse or else you could
damage the instrument. Never replace DSQ II fuses with those of a different type, voltage,
or current rating. Also, to avoid damaging the boards, you should always wear a properly
grounded
electrostatic discharge (ESD) strap on your wrist when replacing fuses.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...