Troubleshooting
Contamination Symptoms
Thermo Scientific
DSQ II Hardware Manual
129
Chemical background due to rhenium oxide (
m/z
185/187, 201/203, 217/219, 233/235,
250/252).
Possible Solutions
These series of rhenium oxide ions come from oxidation of the rhenium filament wire due to
the introduction of air into the ion source while the filament is on. Check for air leaks and
remedy. See
Spectra are observed due to solvents.
Possible Solutions
• There is residual solvent from a cleaning procedure or laboratory background
contamination. When you finish performing a cleaning procedure, allow cleaned
components to dry thoroughly. Warm parts in the GC oven to dry off residual solvent.
• The observed compounds have been introduced through sample injection. Ultimate
source is either a sample solvent or the autosampler rinsing solvent. Optimize GC
method to separate solvent peak from area of interest in the chromatogram.
Table 3.
Solvent Spectra
Solvents
Acetone (
m/z
43, 58, 59)
Hexane (
m/z
41, 43, 56, 57, 58, 85, 86)
Methanol (
m/z
31, 32)
Methylene chloride (
m/z
84/83)
Toluene (
m/z
91, 92)
Trichloroethane (
m/z
151, 153)
Xylene (
m/z
105, 106)
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...