2
Maintaining the DSQ II
Cleaning the Ion Source Components
48
DSQ II Hardware Manual
Thermo Scientific
8. Remove the components surrounding the ion source.
Figure 27.
Removing the Components Surrounding the Ion Source
9. Remove the source spacer, which contains the prefilter entrance lens. Also, compress and
remove the triangular spring clip that the source spacer uses to hold the prefilter lens in
place.
10. Remove the lens stack.
11. Remove the magnet yoke, magnets, and thumbscrews as one unit and set them aside.
12. Disassemble the lens stack.
1
Source Spacer and Prefilter
Lens
6
Two Magnets
2
Lens Stack
7
Magnet Yoke
3
3-Pin Lens Connector
8
Two Thumbscrews
4
Filament
9
Ion Volume (Inside Ion Source)
5
8-Pin Lens Connector
10
Ion Source
2
3
4
5
7
8
1
9
10
6