Troubleshooting
Using Diagnostics
122
DSQ II Hardware Manual
Thermo Scientific
Figure 73.
Diagnostics Flowchart
Running Tune Diagnostics
1. Open the Diagnostics window in
DSQ II Tune
.
a.
Open
Xcalibur
and click
Instrument Setup
.
b. Click the
DSQ II
icon to open the Method Editor.
c.
Click the
Tune
button.
d. Select
Diagnostics | Run Tests
to open the Diagnostics window.
2. Run diagnostics by clicking on the test you want to run.
Did the system fail while you
were running samples?
Did problems occur immediately
after you performed maintenance
on the instrument, data system,
or peripherals?
There may be a hardware
failure. Run diagnostics.
Did problems occur after you changed
the system's configuration, cables, or
peripherals?
The problem is probably mechanical,
not electrical. Verify that alignment,
configurations and cable connections
are correct before you run
Diagnostics
START
Run Diagnostics.
END
No
No
No
Yes
Yes
Yes