4
Vacuum System and Gas Inlets
Vacuum System
Thermo Scientific
DSQ II Hardware Manual
159
Rotary-Vane Pump
The rotary-vane pump is typically located on the floor behind the DSQ II. It is a two-stage
forepump that connects to the outside of the DSQ II. It establishes the vacuum necessary for
properly operating the high vacuum pump.
The rotary-vane pump connects to the high vacuum pump by a piece of 0.75 in. i.d. tubing.
The power cord plugs into the rear of the DSQ II
in the outlet labeled MECH PUMP. This
outlet powers the pump and is controlled by the main circuit breaker and not by the service
switch.
Figure 79.
Rotary-Vane Pump
CAUTION - INSTRUMENT DAMAGE.
Always connect the rotary-vane pump power cord
to the outlet on the rear of the DSQ II
and never into a wall outlet. If the pump is not
plugged into the DSQ II
rear, it will not shut down when the instrument is powered off.
As a result, the pump will continue to run and the DSQ II
will not be able to vent to
atmosphere.
1
Foreline Vacuum Hose
8
Drain Plug
2
Inlet Port
9
MIN (Minimum Level Mark)
3
Gas-Ballast Control
10
Mode Selector
4
Oil Filler Plugs
11
On/Off Switch
5
Outlet Port
12
Handle
6
MAX (Maximum Level Mark)
13
Voltage Indicator
7
Oil-Level Sight-Glass
14
Electrical Inlet Connector
14
3
1
2
5
6
10
9
11
13
12
7
4
8
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...