Maintaining the DSQ II
Maintaining the Rotary-Vane Pump
Thermo Scientific
DSQ II Hardware Manual
65
a.
Remove one of the oil filler plugs from the rotary-vane pump.
b. Add fresh oil into the reservoir until the oil is half way between the MIN and MAX
level marks.
c.
If the oil level goes above the MAX level mark, remove the drain plug and drain the
excess oil into a suitable container.
d. Reattach the oil filler plug.
4. Plug the rotary-vane pump and DSQ II to the power source.
5. Restart the DSQ II.
6. Purge the rotary-vane pump for 10 minutes by using the gas ballast to remove excess gases
from the new oil.
Purging Gas from the Oil in a Rotary-Vane Pump
Purging (or decontaminating) the oil in a rotary-vane pump removes dissolved gases and
low-boiling-point liquids from the oil. Purge the oil if it is cloudy or discolored. You can
purge the oil without interrupting system operation. However, do not purge the pump during
an acquisition or while the electron multiplier or filament is on.
To purge oil from the rotary-vane pump:
1. Make sure you are not in the middle of an acquisition and the electron multiplier and
filament are OFF. Use the
Xcalibur
Tune window to confirm these items are off.
2. Prepare the rotary-vane pump for purging.
CAUTION - INSTRUMENT DAMAGE.
Use only factory-approved rotary-vane pump oil to
avoid damaging the instrument or causing a fire.
Summary of Contents for DSQ II
Page 1: ...DSQ II Hardware Manual 120299 0002 Revision C December 1 2009 ...
Page 4: ......
Page 8: ......
Page 36: ...1 Basic Operations Shutting Down the GC MS System 20 DSQ II Hardware Manual Thermo Scientific ...
Page 198: ...5 Ion Source and Inlet Valve Inlet Valve Upgrade 182 DSQ II Hardware Manual Thermo Scientific ...
Page 202: ...6 Prefilter and Quadrupole 186 DSQ II Hardware Manual Thermo Scientific ...
Page 206: ...7 Ion Detector Electron Multiplier 190 DSQ II Hardware Manual Thermo Scientific ...
Page 242: ...226 DSQ II Hardware Manual Thermo Scientific Index ...