Chapter 4. Troubleshooting and Diagnostics
EK–MOL80–SV. B01
4–21
Table 4–3 Hardware Errors and Recovery Procedures (
continued
)
Error Code
Error Description
Recovery Procedures
(hex.)
4A
Test picker
This may appear when testing for a
cartridge in the picker during an
ISTAT.
•
Replace picker
4B
Switch active picker
Replace picker
4C
Restore picker
Replace picker
4D
Find translate home
Can not translate the picker and/or
sense that it has moved.
Run FIND XLAT HOME test from
the control panel
•
If the picker does not move at
all, check the connections on
the umbilical cable. If the
connections are good and the
picker still does not move,
change the umbilical cable.
•
If the picker moves a little but
does not reach the side of the
frame, the translate motor on
the picker is probably defective.
Change the picker.
•
If the picker moves properly to
the side, the translate sensor is
probably defective. Change the
picker.
4E
Find vertical home
Because a motor test is called
before a find vertical home is
attempted, the vertical motor is
assumed to be at least minimally
functional.
•
Check that the vertical path is
physically clear.
•
Change the vertical motor
(probably a marginal error).