32 · GuardOne
®
· Manual
UV 254 Decon
By pressing UV 254 Decon, you can activate the UV decontamination mode. To automatically stop the UV decontamination after a pre-set
time, you can use the timer function.
NOTICE!
Property damage to plastic materials from prolonged irradiation!
Some types of plastic, such as pipette tips, can become brittle when exposed to prolonged irradiation.
- Do not place plastic containers that are not UV stable into the GuardOne® Workstation.
Operating the UV lamp
1.
To start the UV decontamination, switch the UV status (1) to
.
To stop the UV decontamination, switch the UV status (1) to
.
The UV lamp cannot be switched on while the front door is open.
The UV lamp is switched off automatically as soon as the front door is
being opened. Closing the front door again will not restart the
decontamination procedure.
The Info Expiration Time UV Light Source (2) gives you information
about the remaining life time of the UV lamp.